<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>10065</ErrorNumber>
<ErrorFamily>Hardware</ErrorFamily>
<ErrorMessage>
HARDWARE: DBlower Filter Fault.
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>All</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The debris cabinet is reporting that the filters are not present.  Check the filter
itself, the sensor inside the cabinet, and the cable running to the device.
</text></DescriptionText>
<DescriptionText><text>
If everything looks fine, the error can be masked off temporarily in later versions
of firmware.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
This error should not be reported to the workstation.  If it does, contact the CreoScitex
Response Centre.
</text></DescriptionText>
</OperatorAction>
<Cause>
Filter has been removed from debris cabinet.
</Cause>
<Cause>
Faulty cable from debris cabinet to device.
</Cause>
<Cause>
Faulty sensor inside debris cabinet.
</Cause>
<SeeAlso>
<UnnumberedList>
<Link><LinkAddress>../commands/dr.xml</LinkAddress><LinkTitle>Command: dr</LinkTitle></Link>
<Link><LinkAddress>../commands/mask.xml</LinkAddress><LinkTitle>Command: mask</LinkTitle></Link>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:32:31 2008 PST
</Updated>
<Links>
</Links>
</Error>